Uncertainty Analysis of Newly Developed Low Liquid Flow Calibration Facility at CSIR-NPL, India

Authors

  • Sanjeet Kumar CSIR-National Physical Laboratory
  • Dr. Shiv Kumar Jaiswal CSIR-National Physical Laboratory
  • Dr. Anshul Varshney CSIR-National Physical Laboratory
  • Chatar Singh CSIR-National Physical Laboratory
  • Dr. Shanay Rab University of Brighton
  • Dr. Sanjay Yadav CSIR-National Physical Laboratory

DOI:

https://doi.org/10.56042/ijpap.v62i10.13204

Keywords:

Accuracy, Metrology, Flow measurement, Uncertainty, Liquid flow

Abstract

Over the past decade, micro-liquid flow measurement has become increasingly important, particularly in India, where devices like syringe pumps, infusion pumps, and infusion device analyzers are critical in hospitals and research laboratories. These devices are essential for precise drug dosing in applications such as chemotherapy, pain management, insulin therapy, and other vital treatments, making accurate calibration crucial. To meet this need, the Fluid Flow Metrology Section at the CSIR-National Physical Laboratory (CSIR-NPL) in New Delhi has developed a state-of-the-art calibration facility for micro-liquid flow, capable of measuring flow rates from 10 mL/h to 1500 mL/h. The facility employs a static weighing method, utilizing a dual-range analytical balance with a capacity of 220 g/82 g and a digital timer as reference standards. This setup achieves an expanded uncertainty of 0.30% (at k=2) for both totalized volume and volume flow rate measurements, thereby ensuring the facility's status as a National Metrology Institute (NMI) in micro-liquid flow measurement.

Author Biographies

  • Dr. Shiv Kumar Jaiswal, CSIR-National Physical Laboratory

    Chief Scientist

    Fluid Flow  Metrology Section

    Physico-mechanical Metrology Division

  • Dr. Anshul Varshney, CSIR-National Physical Laboratory

    Scientist

    Fluid Flow Metrology Section

     

  • Chatar Singh, CSIR-National Physical Laboratory

    Technical Officer

    Fluid Flow Metrology Section

     

  • Dr. Shanay Rab, University of Brighton

    Lecturer 

    School of Architecture, Technology, and Engineering 

  • Dr. Sanjay Yadav, CSIR-National Physical Laboratory

    Former Chief Scientist and Head

    Physico-mechanical Metrology Division

Downloads

Published

2024-10-01

How to Cite

Uncertainty Analysis of Newly Developed Low Liquid Flow Calibration Facility at CSIR-NPL, India. (2024). Indian Journal of Pure & Applied Physics (IJPAP), 62(10), 851-863. https://doi.org/10.56042/ijpap.v62i10.13204

Similar Articles

1-10 of 100

You may also start an advanced similarity search for this article.